Article (Scientific journals)
Predictive elution window stretching and shifting as a generic search strategy for automated method development for liquid chromatography
Tyteca, Eva; Liekens, A.; Clicq, D. et al.
2012In Analytical Chemistry, 84 (18), p. 7823-7830
Peer Reviewed verified by ORBi
 

Files


Full Text
ac301331g_pub.pdf
Publisher postprint (2.02 MB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Automated methods; Grid-based; In-silico; Method development; Modeling errors; Parameter spaces; Pharmaceutical industry; Retention properties; Search efficiency; Search strategies; Separation problems; Solution space; Liquid chromatography; Algorithms; Automation; Chromatography, High Pressure Liquid; Drugs, Generic; Pharmaceutical Preparations
Abstract :
[en] We report on the possibilities of a new method development (MD) algorithm that searches the chromatographic parameter space by systematically shifting and stretching the elution window over different parts of the time-axis. In this way, the search automatically focuses on the most promising areas of the solution space. Since only the retention properties of the first and last eluting compounds of the sample need to be (approximately) known, the algorithm can be directly applied to samples with unknown composition, and the proposed solutions are not sensitive to any modeling errors. The search efficiency of the algorithm has been evaluated on an extensive set of random-generated in silico samples covering a broad range of different retention properties. Compared to a pure grid-based search, the algorithm could reduce the number of missed components by 50% and more. The algorithm has also been applied to solve three different real-world separation problems from the pharmaceutical industry. All problems could be successfully solved in a very short time (order of 12 h of instrument time). © 2012 American Chemical Society.
Disciplines :
Chemistry
Author, co-author :
Tyteca, Eva ;  Université de Liège > Agronomie, Bio-ingénierie et Chimie (AgroBioChem) > Analyse, qual. et risques - Labo. de Chimie analytique
Liekens, A.;  Vrije Universiteit Brussel, Department of Chemical Engineering (CHIS-IR), Pleinlaan 2, 1050 Brussels, Belgium
Clicq, D.;  UCB Pharma, Analytical Development Chemicals, Chemin du Foriest, 1420 Braine L'alleud, Belgium
Fanigliulo, A.;  Aptuit, via Fleming 4, 37135 Verona, Italy
Debrus, Benjamin ;  Université de Liège > Département de pharmacie > Chimie analytique
Rudaz, S.;  University of Geneva, University of Lausanne, School of Pharmaceutical Sciences, Boulevard d'Yvoy 20, 1211 Geneva 4, Switzerland
Guillarme, D.;  University of Geneva, University of Lausanne, School of Pharmaceutical Sciences, Boulevard d'Yvoy 20, 1211 Geneva 4, Switzerland
Desmet, G.;  Vrije Universiteit Brussel, Department of Chemical Engineering (CHIS-IR), Pleinlaan 2, 1050 Brussels, Belgium
Title :
Predictive elution window stretching and shifting as a generic search strategy for automated method development for liquid chromatography
Publication date :
2012
Journal title :
Analytical Chemistry
ISSN :
0003-2700
eISSN :
1520-6882
Publisher :
American Chemical Society, Washington, United States - District of Columbia
Volume :
84
Issue :
18
Pages :
7823-7830
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 30 April 2017

Statistics


Number of views
78 (4 by ULiège)
Number of downloads
243 (2 by ULiège)

Scopus citations®
 
26
Scopus citations®
without self-citations
13
OpenCitations
 
24

Bibliography


Similar publications



Contact ORBi