Validation of the Thermodilution Technique for the Estimation of the Cardiac Output in the Unsedated Calf
English
Amory, Hélène mailto [Université de Liège - ULg > Physiology > > >]
Linden, Annick mailto [Université de Liège - ULg > Département des maladies infectieuses et parasitaires > > >]
Desmecht, Daniel mailto [Université de Liège - ULg > Département de morphologie et pathologie > > >]
Rollin, Frédéric mailto [Université de Liège - ULg > Département clinique des animaux de production (DCP) > > >]
Genicot, B. [> > > >]
Lekeux, Pierre mailto [Université de Liège - ULg > Département de sciences fonctionnelles > Physiologie >]
1991
Journal of Veterinary Medicine. A, Physiology, Pathology, Clinical Medicine
Blackwell Publishing
38
6
409-417
Yes (verified by ORBi)
International
0931-184X
1439-0442
Berlin
Germany
[en] thermodilution ; calves
[en] The purpose of this study was to establish a standardized protocol in order to obtain accurate and reproducible cardiac output (CO) measurements by the thermodilution technique in the unsedated calf. In 5 healthy calves, the effect of various temperatures, volumes, rates, means and sites of injection of the thermal indicator were tested for their effect on accuracy and reproducibility of CO measurement by the thermodilution technique, the Fick method being used as the standard technique. Five milliliters of an iced 5% dextrose solution per 100 kg body weight constituted the correct amount of thermal indicator. Manual injection of the indicator bolus and/or injection of this bolus through the proximal lumen of a thermodilution Swan-Ganz catheter gave poor results of CO measurements by the thermodilution technique. On the other hand, results of CO measurements were highly reliable when the bolus injection was performed by an ECG-synchronized injection pump and/or through a right atrial catheter with a larger diameter and a shorter length than the classical Swan-Ganz catheter. It was concluded that an iced 5% solution of dextrose (5 ml/100 kg body weight) injected through a right atrial catheter by an ECG-synchronized pump gives the most accurate and reproducible CO measurements by the thermodilution technique in unsedated calves.
Fonds de la Recherche Scientifique (Communauté française de Belgique) - FNRS & IRSIA Convention n°5131 A
Researchers ; Professionals ; Students
http://hdl.handle.net/2268/6757

File(s) associated to this reference

Fulltext file(s):

FileCommentaryVersionSizeAccess
Restricted access
AmoryH_JVetMed_1991.pdfPublisher postprint2.05 MBRequest copy

Bookmark and Share SFX Query

All documents in ORBi are protected by a user license.