[en] Background: The Accu-Chek Inform glucose monitor is a point-of-care system for testing blood glucose. New test strips, calibrated to deliver glucose plasma-like values, were launched on the market in May 2005. The aim of our study was to perform analytical validation of these new strips. Methods: We compared the new plasma strips with whole blood strips; results for the plasma strips with plasma values obtained using a clinical analyzer and with whole blood values given by the glucose electrode of a blood gas analyzer; and the influence of the type of blood (capillary or venous) on the results obtained by the glucose monitor with the plasma calibrated strips. Results: Plasma strips give on average 7% higher results than the previous whole blood strips. However, the results given by the plasma strips on capillary whole blood, even if well correlated, are not completely comparable with those given by an analyzer for venous plasma. Nevertheless, these plasma strips and the glucose electrode of a blood gas analyzer give comparable results. Conclusions: Accu-Chek Inform plasma strips are a good method for monitoring of blood glucose values in patients with diabetes.
Disciplines :
Laboratory medicine & medical technology
Author, co-author :
Meex, Cécile ; Centre Hospitalier Universitaire de Liège - CHU > Chimie médicale
Poncin, José ; Centre Hospitalier Universitaire de Liège - CHU > Chimie médicale
Chapelle, Jean-Paul ; Université de Liège - ULiège > Département de pharmacie > Chimie médicale
Cavalier, Etienne ; Centre Hospitalier Universitaire de Liège - CHU > Chimie médicale
Language :
English
Title :
Analytical validation of the new plasma calibrated Accu-Chek (R) Test Strips (Roche Diagnostics)
IFCC recommendation on reporting results for blood glucose. Clin Chim Acta 2001;307:205-9.
Kilpatrick ES, Rumley AG, Myint H, Dominiczak MH, Small M. The effect of variations in haematocrit, mean cell volume and red blood cell count on reagent strip tests for glucose. Ann Clin Biochem 1993;30:485-7.
Rao LV, Jakubiak F, Sidwell JS, Winkelman JW, Snyder ML. Accuracy evaluation of a new glucometer with automated hematocrit measurement and correction. Clin Chim Acta 2005;356:178-83.
Stahl M, Brandslund I. Measurement of glucose content in plasma from capillary blood in diagnosis of diabetes mellitus. Scand J Clin Lab Invest 2003;63:431-40.
Chan AY, Swaminathan R, Cockram CS. Effectiveness of sodium fluoride as a preservative of glucose in blood. Clin Chem 1989;35:315-7.