Article (Scientific journals)
Influence of the capillary temperature and the source pressure on the internal energy distribution of electrosprayed ions
Gabelica, Valérie; De Pauw, Edwin; Karas, Michael
2004In International Journal of Mass Spectrometry, 231 (2-3), p. 189-195
Peer Reviewed verified by ORBi
 

Files


Full Text
2004-IJMS-231-189.pdf
Publisher postprint (157.59 kB)
Request a copy
Full Text Parts
2004 IJMS capillary temperature.pdf
Author postprint (164.88 kB)
This is a preprint of an article published in Int J Mass Spectrom © 2004 Elsevier Science B.V.
Download

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
electrospray; internal energy; heated capillary; pressure effect; mass spectrometry
Abstract :
[en] The internal energy distributions in a heated capillary nano-electrospray source have been determined using the "survival yield" method. At low capillary temperatures, the internal energy distribution is characterized by a low-energy tail, which can be attributed to a fraction of ions not fully desolvated in the heated capillary. This low-energy tais is shown to disappear when the source pressure is increased. This explains why increased source pressure is favorable in the case of highly hydrophilic compounds or non-covalent complexes in order to achieve sufficient desolvation without fragmentation. It is also shown that "high temperature-low voltage" are not equivalent to "low temperature-high voltage" source conditions. These observations are important for fundamental issues as well as for source-CID mass spectral library searching applications.
Research center :
CART - Centre Interfacultaire d'Analyse des Résidus en Traces - ULiège
Disciplines :
Chemistry
Author, co-author :
Gabelica, Valérie ;  Université de Liège - ULiège > Département de Chimie (Sciences) > Chimie physique, spectrométrie de masse
De Pauw, Edwin  ;  Université de Liège - ULiège > Département de Chimie (Sciences) > Chimie physique, spectrométrie de masse
Karas, Michael;  Institüt für Pharmazeutische Chemie, Johann-Wolfgang Goethe Universität Frankfurt, Marie-Curie Strasse 9-11, D-60439 Frankfurt am Main, Germany
Language :
English
Title :
Influence of the capillary temperature and the source pressure on the internal energy distribution of electrosprayed ions
Publication date :
2004
Journal title :
International Journal of Mass Spectrometry
ISSN :
1387-3806
Publisher :
Elsevier Science, Amsterdam, Netherlands
Volume :
231
Issue :
2-3
Pages :
189-195
Peer reviewed :
Peer Reviewed verified by ORBi
Funders :
F.R.S.-FNRS - Fonds de la Recherche Scientifique [BE]
AvH - Alexander von Humboldt-Stiftung [DE]
Commentary :
This is a postprint version of an article published in Int J Mass Spectrom © 2004 Elsevier Science B.V.
Available on ORBi :
since 21 July 2008

Statistics


Number of views
147 (2 by ULiège)
Number of downloads
249 (1 by ULiège)

Scopus citations®
 
55
Scopus citations®
without self-citations
52
OpenCitations
 
56

Bibliography


Similar publications



Contact ORBi