Article (Scientific journals)
Automated quantification of creatinine kinase MB isoenzyme in serum by radial partition immunoassay, with use of the Stratus analyzer
Chapelle, Jean-Paul; El Allaf, M.
1990In Clinical Chemistry, 36 (1), p. 99-101
Peer Reviewed verified by ORBi
 

Files


Full Text
Automated quantification of creatine kinase MB... full text.pdf
Publisher postprint (714.96 kB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Abstract :
[en] We evaluated the analytical and clinical performances of a new radial partition immunoassay for measuring the mass concentration of creatine kinase (CK)-MB in serum. All pipetting, washes, incubations and data reduction were performed in 8 min by the Stratus (Dade) fluorometric analyzer. Within-assay and between-assay CVs were respectively 5.5% and 8.4% at 21 micrograms/L, and 4.2% and 3.4% at 48 micrograms/L. Assaying serial dilutions of serum samples with high CK-MB concentrations demonstrated excellent linearity. Results of the Stratus technique correlated well (n = 115, r = 0.98) with those of the Tandem-E CKMB II assay. There was no interference from hemolysis, bilirubin, rheumatoid factor, or added CK-MM (up to 3500 U/L); consequently, CK-MB can be determined in undiluted serum, even in the presence of high total CK activity. The mean CK-MB concentration in 105 blood donors was 1.9 (SD 1.3) micrograms/L. For seven myocardial infarction patients who received prompt fibrinolytic therapy, the mean CK-MB concentration was 4.5 (SD 1.8) micrograms/L at admission, and maximum concentrations, 119 (SD 94) micrograms/L, were recorded 16 h later. CK-MB returned to concentrations less than 10 micrograms/L within 72 h.
Disciplines :
Laboratory medicine & medical technology
Author, co-author :
Chapelle, Jean-Paul ;  Université de Liège - ULiège > Département de pharmacie > Chimie médicale
El Allaf, M.
Language :
English
Title :
Automated quantification of creatinine kinase MB isoenzyme in serum by radial partition immunoassay, with use of the Stratus analyzer
Publication date :
January 1990
Journal title :
Clinical Chemistry
ISSN :
0009-9147
eISSN :
1530-8561
Publisher :
American Association for Clinical Chemistry, Washington, United States - District of Columbia
Volume :
36
Issue :
1
Pages :
99-101
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 22 February 2010

Statistics


Number of views
54 (3 by ULiège)
Number of downloads
122 (0 by ULiège)

Scopus citations®
 
33
Scopus citations®
without self-citations
32
OpenCitations
 
22

Bibliography


Similar publications



Contact ORBi