Complexity of continuous glucose monitoring data in critically ill patients: continuous glucose monitoring devices, sensor locations, and detrended fluctuation analysis methods.
Signal, Matthew; Thomas, Felicity Louise; Shaw, Geoffrey M. et al.
2013In Journal of Diabetes Science and Technology, 7 (6), p. 1492-506
Peer Reviewed verified by ORBi
 

Files


Full Text
Signal et al complexity of CGM in critically ill patients - revised clean.pdf
Author postprint (631.41 kB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Abdomen; Adult; Algorithms; Blood Glucose/analysis; Blood Glucose Self-Monitoring/instrumentation/methods; Calibration; Critical Illness; Female; Fractals; Humans; Hyperglycemia/blood/diagnosis/prevention & control; Intensive Care Units; Male; Middle Aged; Monitoring, Physiologic/instrumentation/methods; Retrospective Studies; Thigh
Abstract :
[en] BACKGROUND: Critically ill patients often experience high levels of insulin resistance and stress-induced hyperglycemia, which may negatively impact outcomes. However, evidence surrounding the causes of negative outcomes remains inconclusive. Continuous glucose monitoring (CGM) devices allow researchers to investigate glucose complexity, using detrended fluctuation analysis (DFA), to determine whether it is associated with negative outcomes. The aim of this study was to investigate the effects of CGM device type/calibration and CGM sensor location on results from DFA. METHODS: This study uses CGM data from critically ill patients who were each monitored concurrently using Medtronic iPro2s on the thigh and abdomen and a Medtronic Guardian REAL-Time on the abdomen. This allowed interdevice/calibration type and intersensor site variation to be assessed. Detrended fluctuation analysis is a technique that has previously been used to determine the complexity of CGM data in critically ill patients. Two variants of DFA, monofractal and multifractal, were used to assess the complexity of sensor glucose data as well as the precalibration raw sensor current. Monofractal DFA produces a scaling exponent (H), where H is inversely related to complexity. The results of multifractal DFA are presented graphically by the multifractal spectrum. RESULTS: From the 10 patients recruited, 26 CGM devices produced data suitable for analysis. The values of H from abdominal iPro2 data were 0.10 (0.03-0.20) higher than those from Guardian REAL-Time data, indicating consistently lower complexities in iPro2 data. However, repeating the analysis on the raw sensor current showed little or no difference in complexity. Sensor site had little effect on the scaling exponents in this data set. Finally, multifractal DFA revealed no significant associations between the multifractal spectrums and CGM device type/calibration or sensor location. CONCLUSIONS: Monofractal DFA results are dependent on the device/calibration used to obtain CGM data, but sensor location has little impact. Future studies of glucose complexity should consider the findings presented here when designing their investigations.
Disciplines :
Engineering, computing & technology: Multidisciplinary, general & others
Endocrinology, metabolism & nutrition
Author, co-author :
Signal, Matthew
Thomas, Felicity Louise ;  Université de Liège - ULiège > Doct. sc. ingé. & techno. (électr., électro. & inf- paysage)
Shaw, Geoffrey M.
Chase, J. Geoffrey
Language :
English
Title :
Complexity of continuous glucose monitoring data in critically ill patients: continuous glucose monitoring devices, sensor locations, and detrended fluctuation analysis methods.
Publication date :
2013
Journal title :
Journal of Diabetes Science and Technology
ISSN :
1932-3107
eISSN :
1932-2968
Publisher :
Diabetes Technology Society, United States - California
Volume :
7
Issue :
6
Pages :
1492-506
Peer reviewed :
Peer Reviewed verified by ORBi
Commentary :
(c) 2013 Diabetes Technology Society.
Available on ORBi :
since 10 May 2016

Statistics


Number of views
43 (4 by ULiège)
Number of downloads
167 (0 by ULiège)

Scopus citations®
 
13
Scopus citations®
without self-citations
11
OpenCitations
 
13

Bibliography


Similar publications



Contact ORBi