Article (Scientific journals)
Coil on the loop and selectively enhanced stiffness for improved control: feasibility study in the neonatal ovine model of the large patent ductus arteriosus.
Grabitz, R. G.; Freudenthal, F.; Sigler, M. et al.
1997In Investigative Radiology, 32 (10), p. 636-43
Peer Reviewed verified by ORBi
 

Files


Full Text
Coil on the loop and selectively enhanced stiffness for improved control.pdf
Publisher postprint (3.88 MB)
Request a copy

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Animals; Animals, Newborn; Cardiology/instrumentation; Ductus Arteriosus, Patent/therapy; Equipment Design; Sheep
Abstract :
[en] RATIONALE AND OBJECTIVES: Several devices have been suggested and tested for interventional closure of the persistent ductus arteriosus. Coils were found effective only in small ducts due to their lack of maneuverability and physical limitation of grip forces leading to risk of embolization. The authors evaluated the feasibility to place single coils with selectively enhanced stiffness into high shunting ductus arteriosus, the coils being deployed and controlled through a veno-arterial loop in a bovine model. METHODS: "Double-cone" shaped, stainless steel coils with enhanced stiffness of the outer rings were mounted on either end on a nitinol core wire. A snap-in mechanism attaches the coil to this delivery wire and is freed by a pusher system of coiled steel wire that is advanced over the core wire. Forming a veno-arterial loop across the patent ductus allows for repositioning into the pulmonary artery or aortic catheter. A chronic lamb model of large patent ductus arteriosus (PDAs) (> or = 9 mm) was used in which ductus patency was secured by a protocol of repetitive angioplasties. Different systems (n = 10) were placed having retrieved the previous coil by a snare after definitive release. RESULTS: Placement of coils was possible in all 10 attempts. The coils were pulled back into the catheters between five and eight times before definitive release. CONCLUSIONS: The device allows controlled placement of single coils in our model of large PDAs and may lead the way to overcome previous limitations of coils. Clinical trials are warranted.
Disciplines :
Cardiovascular & respiratory systems
Pediatrics
Author, co-author :
Grabitz, R. G.;  Rheinisch - Westfälische Technische Hochschule Aachen - RWTH > Department of Pediatric Cardiology and Interdisciplinary Center of Clinical Research on Biomaterials
Freudenthal, F.;  rwth > Pediatric Cardiology
Sigler, M.;  Rheinisch - Westfälische Technische Hochschule Aachen - RWTH > Department of Pediatric Cardiology and Interdisciplinary Center of Clinical Research on Biomaterials
SEGHAYE, Marie-Christine ;  Rheinisch - Westfälische Technische Hochschule Aachen - RWTH > Pediatric Cardiology
Boosfeld, C.;  Consultant to pfm Produkte für die Medizin GmbH, Cologne, Germany
von Bernuth, G.;  Rheinisch - Westfälische Technische Hochschule Aachen - RWTH > Pediatric Cardiology
Language :
English
Title :
Coil on the loop and selectively enhanced stiffness for improved control: feasibility study in the neonatal ovine model of the large patent ductus arteriosus.
Publication date :
October 1997
Journal title :
Investigative Radiology
ISSN :
0020-9996
eISSN :
1536-0210
Publisher :
Lippincott Williams & Wilkins, Hagerstown, United States - Maryland
Volume :
32
Issue :
10
Pages :
636-43
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 07 February 2014

Statistics


Number of views
47 (0 by ULiège)
Number of downloads
0 (0 by ULiège)

Scopus citations®
 
5
Scopus citations®
without self-citations
3
OpenCitations
 
4

Bibliography


Similar publications



Contact ORBi