A pilot study of the safety and performance of the Otologics fully implantable hearing device: transducing sounds via the round window membrane to the inner ear.
[en] OBJECTIVES: The safety and performance of the Otologics fully implantable hearing device were assessed in adult patients with mixed conductive and sensorineural hearing loss. METHODS: The subcutaneous microphone of this fully implantable device picks up ambient sounds, converts them into an electrical signal, amplifies the signal according to the user's needs, and sends it to an electromechanical transducer. The transducer tip is customized with a prosthesis in order to be in contact with the round window membrane and is protected by fascia; this translates the electrical signal into a mechanical motion that directly stimulates the round window membrane and enables the user to perceive sound. The implanted battery is recharged daily via an external charger and the user can turn the implant on and off as well as adjust the volume with a hand-held remote control. In this pilot study, 6 patients with mixed conductive and sensorineural hearing loss were implanted with the Otologics fully implantable hearing device. Pre- and postoperative air conduction, bone conduction, as well as aided and unaided thresholds and speech scores were measured. RESULTS: No significant differences between preoperative and postoperative pure-tone averages were noted. Average improvement ranged from 19.16 to 35.8 dB of functional gain across audiometric frequencies with a mean of 26.17 +/- 5.15 dB. Long-term average functional gain at 12 months was 20.83 +/- 6.22 dB. Word recognition scores demonstrated significant differences between unaided and implant-aided conditions. CONCLUSIONS: Preliminary results of this trial of the Otologics fully implantable hearing device provide evidence that this fully implantable device is capable of efficiently transferring the sound to the inner ear via the round window membrane in patients with mixed hearing loss.
Disciplines :
Otolaryngology Gastroenterology & hepatology
Author, co-author :
Lefèbvre, Philippe ; Université de Liège - ULiège > Département des sciences cliniques > Oto-rhino-laryngologie et audiophonologie
Martin, Christine ; Centre Hospitalier Universitaire de Liège - CHU > HOSPITALISATION - GASTRO & NEPHROLOGIE (SN +4B)
Dubreuil, C.
Decat, M.
Yazbeck, A.
Kasic, J.
Tringali, S.
Language :
English
Title :
A pilot study of the safety and performance of the Otologics fully implantable hearing device: transducing sounds via the round window membrane to the inner ear.
Colletti V, Soli SD, Carner M, Colletti L: Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window. Int J Audiol 2006; 45: 600-608.
Cox RM, Alexander GC: The abbreviated profile of hearing aid benefit. Ear Hear 1995; 16: 176-186.
Dumon T, Zennaro O, Aran JM, Bebear JP: Piezoelectric middle ear implant preserving the ossicular chain. Otolaryngol Clin North Am 1995; 28: 173-187.
Fisch U, Cremers CW, Lenarz T, Weber B, Babighian G, Uziel AS, Proops DW, O'Connor AF, Charachon R, Helms J, Fraysse B: Clinical experience with the Vibrant Soundbridge implant device. Otol Neurotol 2001; 22: 962-972.
Jenkins HA, Niparko JK, Slattery WH, Neely JG, Fredrickson JM: Otologics Middle Ear Transducer Ossicular Stimulator: performance results with varying degrees of sensorineural hearing loss. Acta Otolaryngol 2004; 124: 391-394.
Kasic JF, Fredrickson JM: The Otologics MET ossicular stimulator. Otolaryngol Clin North Am 2001; 34: 501-513.
Ko WH, Zhu WL, Maniglia AJ: Engineering principles of mechanical stimulation of the middle ear. Otolaryngol Clin North Am 1995; 28: 29-41.
Maniglia AJ, Ko WH, Garverick SL, Abbass H, Kane M, Rosenbaum M, Murray G: Semiimplantable middle ear electromagnetic hearing device for sensorineural hearing loss. Ear Nose Throat J 1997; 76: 333-338, 340-341.
Siegert R, Mattheis S, Kasic J: Fully implantable hearing aids in patients with congenital auricular atresia. Laryngoscope 2007; 117: 336-340.
Snik AF, Mylanus EA, Cremers CW, Dillier N, Fisch U, Gnadeberg D, Lenarz T, Mazolli M, Babighian G, Uziel AS, Cooper HR, O'Connor AF, Fraysse B, Charachon R, Shehata-Dieler WE: Multicenter audiometric results with the Vibrant Soundbridge, a semi-implantable hearing device for sensorineural hearing impairment. Otolaryngol Clin North Am 2001; 34: 373-388.
Venail F, Lavieille JP, Meller R, Deveze A, Tardivet L, Magnan J: New perspectives for middle ear implants: first results in otosclerosis with mixed hearing loss. Laryngoscope 2007; 117: 552-555.
Wever EG, Lawrence M: The transmission properties of the middle ear. 1950. Ann Otol Rhinol Laryngol 1992; 101: 191-204.
Zenner HP, Limberger A, Baumann JW, Reischl G, Zalaman IM, Mauz PS, Sweetow RW, Plinkert PK, Zimmermann R, Baumann I, De Maddalena H, Leysieffer H, Maassen MM: Phase III results with a totally implantable piezoelectric middle ear implant: speech audiometry, spatial hearing and psychosocial adjustment. Acta Otolaryngol 2004; 124: 155-164.