Abstract :
[en] Introduction : the purpose of this study was to investigate the validity and reliability of an inertial dynamometer using accelerometry.
Methods : fourteen subjects were tested during three successive sessions at four increasing loads (30, 50, 70 and 95% of the 1RM). Maximal force, velocity and power performances were simultaneously assessed by the Myotest (Myotest S.A., Switzerland), an inertial dynamometer using accelerometry, and by another control valid dynamometer witch combine accelerometer and linear position transducer. At three first loads, only minor differences appeared between both dynamometers, correlations were excellent and coefficients of variation were good and similars. However, at 95% of the 1RM, Myotest validity and reliability became insufficient.
Conclusion : Bench press inertial assessment with accelerometer appeared valid and reliable at 30, 50 and 70% of the 1RM, but not at 95% of the 1RM.
Scopus citations®
without self-citations
21