Article (Scientific journals)
Evaluation of some parameters influencing the drug delivery from a dry powder inhalation device using an in vitro model of the horse airways
Duvivier, Dominique Hannia; Chiap, Patrice; Crommen, Jacques et al.
1997In Veterinary Research, 28 (6), p. 557-564
Peer Reviewed verified by ORBi
 

Files


Full Text
DuvivierDH_VetRes_1997.pdf
Publisher postprint (2.42 MB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Dry powder inhalation; ipratropium bromide; horse; drug delivery
Abstract :
[en] The aim of this study was to determine the effect of breathing pattern, air humidity and position of the device on the delivery of an aerosol generated by a dry powder inhalation (DPI) device (Inhalator M). The in vitro inhalation study was performed using the cascade impaction method (Andersen Sampler) adapted to imitate nasal breathing. The amount of ipratropium found in the device, the artificial upper airways and the six stages of the Andersen Sampler was measured using high precision liquid chromatography. Stage 1 of the Andersen Sampler was considered to be the respirable fraction and stages 2 to 6 to be the non-respirable fraction. It was concluded that the theoretical respirable fraction of ipratropium obtained after DPI through Inhalator M was influenced by relative air humidity, air flow and the position of the device, whereas the number of successive inspirations and the duration of inspiration did not affect this fraction of the drug.
Disciplines :
Veterinary medicine & animal health
Anatomy (cytology, histology, embryology...) & physiology
Author, co-author :
Duvivier, Dominique Hannia;  Université de Liège - ULiège > Faculté de médecine vétérinaire > Physiologie
Chiap, Patrice ;  Université de Liège - ULiège > Département de pharmacie > Analyse des médicaments
Crommen, Jacques ;  Université de Liège - ULiège > Département de pharmacie > Analyse des médicaments
Lekeux, Pierre ;  Université de Liège - ULiège > Département de sciences fonctionnelles > Physiologie
Language :
English
Title :
Evaluation of some parameters influencing the drug delivery from a dry powder inhalation device using an in vitro model of the horse airways
Publication date :
1997
Journal title :
Veterinary Research
ISSN :
0928-4249
eISSN :
1297-9716
Publisher :
EDP Sciences, Les Ulis, France
Volume :
28
Issue :
6
Pages :
557-564
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 09 March 2012

Statistics


Number of views
57 (2 by ULiège)
Number of downloads
73 (1 by ULiège)

Scopus citations®
 
5
Scopus citations®
without self-citations
4

Bibliography


Similar publications



Contact ORBi